Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(4)

April 13, 2021

Latest company case about Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(4)

 

Fault phenomenon


In the Review screen, a Red Cross will appear on the right side of the patient information. And a corresponding Red Cross will appear on the lower right side of the open image.



Troubleshooting


Is caused by data file corruption, the red check will not be output, but there is no red image in the patient's check can be output alone.Call the background Reset Patient data-base program to resolve the fault.