Detail Information |
|||
Category: | Ultrasound And Doppler | Description: | Sonography Instrument |
---|---|---|---|
System: | Ultrasound Machine Manufacturers | Warranty: | 60 Days After Fixed |
Condition: | GE Voluson E9 Ultrasound Maintenance | Purpose: | Probe Frequency Ultrasound |
Highlight: | GE Logiq E9 Ultrasound Machine Repair,5161631Ultrasound Machine Repair,Medical Imaging Ultrasound Machine Repair |
Product Description
Failure phenomenon
GE LOGIQ E9 color Doppler ultrasound frequently crashes after turning on.
This fault is a startup fault. First, it is necessary to determine whether the fault is at the front or back end of the system.
Open the BEP shell, turn it on, and find that the rear power indicator shows the following: PS-ON and 5 V stdby LED indicators are on, other indicators are off. It can be seen that +5 V STB has output and PS-ON has action, but there is no +48 V voltage directly output by the main power supply. The reasons for this phenomenon may be:
1. The main power supply fails;
2. The +48 V load has an overcurrent short-circuit phenomenon.
After disconnecting the +48 V load, the fault remains. It is judged that the fault is caused by poor main power supply. For further confirmation, after the main power supply was removed and tested separately, it was found that the main power supply was damaged and there was no +48 V output. After replacement, the fault was eliminated.
Repair List of Featured Ultrasonic Board & Assy | |||
Brand | System | Description | Part Number |
GE | Logiq P5 | CPU mainboard | 5168431 |
GE | Logiq P6 | CPU mainboard | |
GE | Logiq E9/vivid E9 | GFI2 | 5161631 |
GE | Voluson E6/Voluson E8 | RSR | KTI301394-2/KTI196357 |
GE | Voluson E6/Voluson E8 | RFI/ RFI21b Board | KTI300614/KTI302197-6 |
GE | Voluson S6/S8/P8 | BF64 | 5396937-2 |
GE | Voluson S6/S8/P8 | RFS board | 5364098-2/5364098-3 |
GE | Voluson S6/S10/P8 | BF192 board | 5357234 |
Aloka | Alpha 5 | mainboard | EP493700DD |
Aloka | Alpha 6 | RX Beam Former | EP555501AA |
Aloka | Alpha 6 | CW board | EP555700 |
Aloka | Alpha 6 | CPU mainboard | EP537000/EP563000 |
Aloka | Alpha 6 | TX/RX | EP555300 |
Aloka | Alpha 6/Alpha 7 | DOM | EP550000AA |
Aloka | Alpha 10 | RX Beamformer board | EP531800CE |
Aloka | Alpha 10 | wave generator Board | EP495000HH/EP495000HJ/EP495000FF |
Hitachi | F37 | TX/RS | EP557400/EP557300 |
Hitachi | F37 | mainboard | EP575700BC/EP560800 |
Aloka | F75 | USP board | EP556600 |
Aloka | F75 | Cell Board | EP558800 |
Hitachi | HI VISION Avius | RDBF board | 7513907A/7513612 |
Hitachi | Arietta 60 | CELL | EU-9160C |
Hitachi | Arietta 70 | TX | EP572300AA |
Hitachi | Arietta 70 | CELL | |
Hitachi | Noblus | CELL | |
Hitachi | Noblus | CONT-B | |
Hitachi | Ascendus | CELL | 7352830A |
Siemens | X150/G40 | BE | 10131804 |
Siemens | X300 | BE 2.0/BE 3.0/BE 7.5 | 10131990 |
Siemens | X700 | BE | 10136465 |
Siemens | X700 | RC | 10136467 |
Siemens | X700 | TI | 11014292 |
Siemens | Antares | BE | 7472421 |
Siemens | Antares | CB | 7288504 |
Siemens | S2000 | RM300 | 10852163 |
Siemens | S2000 high version/S3000 | RM301 | 10854749 |
iU22/iE33 | EMB excelsior motherboard | 453561419431/453561382231/453561419431 | |
iU22/iE33 | UMB Unified Motherboard | 453561254492/453561360251 | |
HD15 | ACB acquisition control board | 453561197145/453561197144 | |
HD15 | AIM board | 4.53561E+11 | |
CX50 | control panel | 4.53562E+11 | |
CX50 | mainboard | 453561368033/453561622163/453561495832 | |
Affiniti 70 | ACQ acquisition control board | ||
EPIQ 5/EPIQ 7 | ACQ Acquisition Module board | 453561734844/453561704246 | |
Biosound | Mylab Series | CPU mainboard | 9501361000 |
Toshiba | Aplio 300/400/500 | mainboard | PM30-38696 |
Toshiba | Aplio 500 | RX | ywm2056*a |
Toshiba | Xario 200 | TX | PM30-32732-1 |
Toshiba | Xario 200 | RX | PM30-32733-1 |
Remarks | |||
1.The repair cycle is within 3-5 days. | |||
2.Warranty after fixed:60 days. | |||
3. Kindly send us the estimated main faults (better to have image) for our engineer to prejudge. | |||
4.Package price is available if repair 3 pcs or more at one time. | |||
5.Extra discount for VIP partners. | |||
6.You could have different part No.under same description. |
Enter Your Message